Karl Storz XENON NOVA 175 Gebrauchsanweisung Seite 44

Kaltlicht-fontäne
Vorschau ausblenden Andere Handbücher für XENON NOVA 175:
Inhaltsverzeichnis

Werbung

36
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Guidance and manufacturer's declaration – electromagnetic immunity
XENON NOVA
®
175 model 20 1315 20 is intended for use in the electromagnetic environment specified below.
The customer or user of the XENON NOVA
Immunity test
EN/IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
IEC 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short
interruptions and voltage
<5% U
variations on power supply
(>95% dip in U
input lines
for 0.5 cycle
IEC 61000-4-11
40% U
(60% dip in U
for 5 cycles
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
Appendix
Electromagnetic Compatibility
(EMC) Information
Table 202
®
175 model 20 1315 20 should ensure that it is used in such an environment.
Compliance level
Complies
± 6 kV contact
± 8 kV air
Complies
± 2 kV for power supply lines
± 1 kV for input/output lines
Complies
± 1 kV differential mode
± 2 kV common mode
Complies
*
<5% U
*
T
T
)
(>95% dip in U
T
for 0.5 cycle
Complies
40% U
T
T
)
(60% dip in U
)
T
T
for 5 cycles
Complies
70% U
T
T
)
(30% dip in U
)
T
T
for 25 cycles
Complies
<5% U
T
T
)
(>95% dip in U
T
for 5 sec
Complies
3 A/m
Anexo
Indicaciones sobre compatibilidad
elec tromagnética (CEM)
Electromagnetic environment - guidelines
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
)
T
If the user of the equipment or system requires
continued operation during power mains interruptions,
it is recommended that the equipment or system be
powered from an uninterruptible power supply.
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.