Invacare® HomeFill® II

7.3 Typical Product Parameters

Regulatory Listing:
Operating Environmental
Conditions:
Atmospheric Pressure:
Storage and Transport:
Input Pressure Required:
Input Flow Required:
Input O
% Required:
2
Width:
Height:
Depth:
Weight:
Shipping Weight:
Max Applied Part
Temperature:
28
Downloaded from
www.Manualslib.com
EN/IEC 60601–1
EN/IEC 60601–1–2
41 to 104° F (5 to 40° C) at
15 to 60% non-condensing
humidity
700 to 1060 hPa
-25 to 158° F (-23 to 70° C)
at 15 to 95% non-condensing
humidity
14 - 21 PSI (96.5-144.7 kPa)
2 L/min
>90% O
2
20 ¼ inches (51.44 cm)
15 inches (38.1 cm)
16 inches (40.64 cm)
33 lbs (14.97 kg)
39 lbs (17.69 kg)
122° F (50° C)
manuals search engine
Expected Service Life:
Electrical Ratings:
The expected service life of
this product is 5 years or
5000 hours of operation when
used in accordance with the
safety instructions, maintenance
intervals and correct use, stated
in this manual. The effective
service life can vary according
to frequency and intensity of
use. Maintenance should be
done at least every week. Refer
to Maintenance section.
230V~, 1.0A, 50Hz