Invacare® HomeFill® II
PROBLEM
Fill times exceed maximum recommended
fill times or GREEN filling light stays On.
Excessive noise when compressor is On.
Hissing sound from Regulator when flow
selector is Off.
Regulator is On but O
delivered.
26
Downloaded from
www.Manualslib.com
SOLUTION
1. Ensure the cylinder regulator is set to Off.
2. Check the recommended fill times shown in 7.2 Cylinder Fill Times, page 27.
3. Check that the cylinder is connected properly. Turn the compressor Off (O). Remove the
cylinder and reinstall. Turn the compressor On ( ).
4. Check the compressor/concentrator line for proper connection and/or damage (i.e. - leaks
or kinks).
5. Check the concentrator liter flow setting is at 3 L/min. or less. If necessary, adjust the liter
flow setting (refer to the concentrator user manual). If the GREEN filling light does not go Off,
the compressor and/or concentrator may need service. Contact your home health care provider
or Invacare.
1. Call home health care provider or Invacare for service.
1. Turn flow selector Off.
2. If hissing sounds continue, call home health care provider or Invacare.
is not being
1. Refill cylinder
2
2. Check that cannula is straight and not pinched. If damaged, replace.
3. Call home health care provider or Invacare for service.