Herunterladen Inhalt Inhalt Diese Seite drucken

Kangaroo Joey Bedienungsanleitung Seite 30

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 95
Caregiver Alarm Notification
All alarms are intended to be heard by operators that are within hearing range of the pump buzzer. The pump buzzer is located on
the back of the pump. The pump is designed so the alarm can be heard within the patient's room, at a minimum. The display and
LED alarm indicators are intended to be seen by an operator within the room, facing the front of the pump. Since audible alarms are
limited by distance, it is recommended that the operator conduct a check to determine at what distance the alarm can still be heard.
Note: Going outside of the patient's room may make hearing the alarms more difficult.
The Error, Warning, and Information Screens are described below:
System Error (High Priority Alarm)
The system error screen is the most general form of error, Figure 24. The only way to exit from a
System Error is to power down. No CONTINUE option is allowed because of the severity of the
error. An error number is also displayed on the screen, for reference purposes. This number should
be reported when calling the Customer Service Support Line. Detection of this alarm condition
may take up to 1 minute to occur under normal operating conditions.
Error # Related Sub-system
0
Internal timer stopped.
1
Flash Chip missing OR version incompatible.
2
Too many graphic screens in memory.
4
Never got a ADC ISR for motor current.
5
Never got a ADC ISR for battery level.
6
Never got a ADC ISR for battery temperature.
7
Never got a ADC ISR for Ultrasonic Upstream
Occlusion or Bag-Empty checks.
8
Never got a ADC ISR for Ultrasonic during
Downstream Occlusion.
10
Flash Write Error.
11
Flash Read Error.
12
General Stopcock Time Out.
Turn the power off and back on to try to clear the error. If the error cannot be resolved in a timely manner, press
operation of the pump and put a different pump into service.
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
Error # Related Sub-system
13
Time Out for Buzzer Test.
16
Flash Screen Read Error.
18
Battery Unplugged or Other Error.
19
Flash Languages Read Error.
20
Could not find Stop Cock Flush Position during
prime routine.
21
Could not find Stop Cock Feed Position during
prime routine.
22
Motor Error during Tubing AutoLoad.
23
Motor Failsafe Time Out Error. Possible damaged
gearbox encoder or flex cable.
25
Display Lockup Time Out Error.
33
Motor Control Failure.
26
Figure 24. General System
Error screen. See Customer
Service section.
POWER to stop
Table of Contents

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Kangaroo Joey

Inhaltsverzeichnis