Parameter input
Parameter
Parameter
Channel
A:
Select the
channel.
Transducer
Connected Transd
Select the
transducer.
Outer diameter
62.00
mm
Enter the
outer pipe diameter.
Pipe material
Carbon Steel
Select the
pipe material.
Wall thickness
3.00
mm
Enter the
pipe wall thickness.
Lining
>NO<
YES
Is pipe
lining present?
Roughness
Auto
Enter
the roughness.
Fluid
Water
Select
the fluid.
Fluid temp.
22.0
C
Enter the
fluid temperature.
Extension cable
0.0
m
Enter the length
of the extension cable.
Start measurement
Measurement
Channel
>A< B Y Z
- - -
x
Measurement
Select the
channel.
A: Meas. point no.
A
Enter the
measuring point number.
A: Sound path
2
NUM
Confirm
the value.
Transd. distance
54
mm
Use coupling foil or apply
a bead of acoustic
coupling compound to
the contact surface
of the transducers.
Mount the transducers on the
pipe at the recommended
transducer distance a.
A: Measurement
Press
or
until bar
graph Q is displayed in the lower
line.
Change transducer position slightly,
until bar graph Q has max. length.
Notes on transducer mounting
• Observe the recommended distance between
the measuring point and the disturbance point.
• Clean the pipe.
• Use coupling foil or apply a bead of acoustic
coupling compound to the contact surface
of the transducers.
• Mount the transducers laterally to the pipe,
if possible.
Transd. distance(50.00)
54
mm
Measure and enter trans-
ducer distance a.
A:Volum. flow rate
54.5
m3/h
Measurement
QIFLUXUS_FG72xV1-0-2, 2019-10-10
Navigation
• vertical selection (
): key
and
&
• horizontal selection (
): key
"
• return to main menu: key BRK
• delete: key C
Channel
x
: channel is activated
-
: channel is deactivated
: no parameters
• select the channel: key
and
"
• activate the channel: key
and
&
Input of sound path
• sound path (even number):
The transducers are mounted on
the same side of the pipe.
• sound path (uneven number):
The transducers are mounted on
opposite sides of the pipe.
Note!
When the transducers are mounted
correctly, the engravings on the trans-
ducers form an arrow.
Measurement
Measurement
Press
.
Stop measurement
and
$
$
a
a
a - transducer distance
equipotential
bonding ter-
minal
Show parameters
Show measurement