Attention!
Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Transmitter
A
display
transducers
channel B
transducers
channel A
X_AV
X_AR
X_BV
X_BR
X2
X3
sensor
sensor
module
module
channel A
channel B
Connection of Transducers
Attention!
For good high frequency shielding, it is important to
ensure good electrical contact between the external
shield and the cap nut (and the housing).
Connection to transmitter
terminal
extension cable
AV
white or marked cable (core)
AVS
white or marked cable (internal shield)
ARS
brown cable (internal shield)
AR
brown cable (core)
terminal
transducer cable with stripped cable ends
AV
transducer
(brown cable, marked white)
AVS
transducer
(red cable)
ARS
transducer
(red cable)
AR
transducer
(brown cable)
terminal
transducer cable with SMB connectors
X_AV
SMB connector (brown cable, marked white)
X_AR
SMB connector (brown cable, marked black)
Connection to junction box
terminal
extension cable
TV
white or marked cable (core)
TVS
white or marked cable (internal shield)
TRS
brown cable (internal shield)
TR
brown cable (core)
terminal
transducer cable with stripped cable ends
V
transducer
(brown cable, marked white)
VS
transducer
(red cable)
RS
transducer
(red cable)
R
transducer
(brown cable)
terminal
transducer cable with SMB connectors
XV
SMB connector (brown cable, marked white)
XR
SMB connector (brown cable, marked black)
Q
keyboard
B
NEXT
ON
DISP
SNAP
Q-
Q+
MUX
DISP
Q
OFF
FAST
MODE
Ethernet interface
LAN
USB
USB interface
power supply
outputs
inputs
connection system TS
junction box JB01
Connection of Power Supply
terminal
PE
N(-)
L(+)
The voltage is indicated on the plate below the terminal strip.
equipotential
bonding terminal
transmitter
transducer, junction box and transmitter have the same electrical potential
connection AC
earth
neutral
phase
Quick Start Guide
FLUXUS *72*
QIFLUXUS_FG72xV1-0-2, 2019-10-10
Sensor module
The supplied sensor mod-
ule (SENSPROM) has to
be connected to the corre-
sponding terminals of the
transmitter. The serial
numbers of the sensor
module and the transducer
must be identical.
junction box JB02, JB03
transmitter
connection DC
earth
-DC
+DC
EN