Weight display, reset and autotare
On power up, the display shows the current net weight.
Switching between net weight and gross weight display
4

2
3
5
Switching between numeric display / bar-graph of the gross weight
4

2
3
5
Restore the zero (semi-automatic zero) in the gross weight display
4

3 sec.
2
3
5
If there is an autotare, this is automatically canceled. The semi-automatic zero operation is canceled when
the instrument is turned off.
Auto weighted tare (Autotara) in net weight display
4

3 sec.
2
3
5
If the autotare with gross weight = 0 is performed, the possible tare value is canceled. The entered tare val-
ue is not saved when the instrument is turned off. If a tare is inserted, this is signaled by the second LED on
(on: tare inserted).
Press the
key to switch the display from net weight to gross weight and
vice versa. The displayed value is indicated by the upper LED (lit: net
1
weight). If the tare is not entered, the net weight is equal to the gross weight.
5
This operation is saved on shutdown. In case of negative weight, the minus
sign is displayed before the most significant digit. In the case of a negative
0
weight greater than 9999, the minus sign is displayed alternating with the
5
most significant digit.
Press the
key to switch from the numerical display of the weight to the
graphic representation of the gross weight and vice versa. The resolution is
=
limited to 15 divisions and therefore each segment represents 1/15 of the full
=
scale. The bar-graph display is indicated by both flashing LEDs.
=
This operation is performed to correct small zero movements of the scale.
To carry out the zeroing function it is necessary to switch the display to the
gross weight (NET LED off). The gross weight reset command is not per-
=
formed in the following situations:
=
Unstable weight (the weight does not stabilize within 3 seconds from the
=
reset command)
Gross weight, greater than the original zero calibration, greater (positive
or negative) than the set 0bAnd value. In this case it is necessary to
enter the calibration menu and perform the calibration
To carry out the autotare function it is necessary to switch the display to the
net weight. (NET led on). The autotare command is not performed in the
following conditions:
=
Unstable weight (weight does not stabilize within 3 seconds)
=
Gross weight negative
=
Gross weight exceeds the maximum capacity of the system