Functional description
GB
!
Zero
Test
!
Zero
Test
Note:
The description of the methods and reagent handling is not included in this manual; it is
contained in a separatedocument.
Store
Date
1
diS
Time
Cal
On
Off
!
Mode
Zero
Test
!
MD1X0/200_1b 9b/2020

Countdown / reaction period

For methods with a reaction time, a countdown function can be swit-
ched on during the test for some methods.
press and hold
press
release
Countdown/reaction time runs, measurementtakesplace automatically
after the time runs out.
Interrupt countdown/reaction time
Attention:
Non-compliance with reaction periods leads to incorrect test
results.

Recall of stored data

The device has a ring memory. The last 16 measured values can be
retrieved.
Switch the unit on
pressfor more than 4 seconds, then release,to accessthe recall menu
Scroll through data sets
Repeat the data set
Return to measurement mode
Displayin the following format (automatically proceedsevery3 seconds
until result is displayed):
Number n xx (xx: 16...1)
Year
YYYY (e.g. 2014)
Date
mm.dd (monthmonth:dayday)
Time
hh:mm (hourhour:minuteminute)
T est
Method
Result
x.xx