Adjusting the Diopter (fig. 19)
1. Turn the eyepiece 1 to adjust the diopter while
looking through it.
• The diopter range of the eyepiece is ±5 di-
opter. The number aligned to the line on the
viewing head is the diopter in use.
Adjusting the tension adjustment collar (fig.
20)
Turn the tension adjustment collar 1 with your
fingers. When the collar is turned in the direction
of the arrow, the tension of the coarse adjust-
ment knob 2 increases. Turning the collar in the
opposite direction decreases the tension.
If the nosepiece descends on its own or if the
specimen gets out of focus quickly even when it
is brought into focus using the fine adjustment
knob 3, it means the tension of the coarse ad-
justment knob is too low. Turn the collar in the
direction of the arrow to increase the tension.

4. Technical Specifications

4.1 Main Specifications

Optical System
Viewing Head
Eyepiece
Nosepiece
Objective
Focusing
Condenser
Stage
Illumination
ENGLISH
Limited optical system/Infinite optical system
Seidentopf Binocular Head, 30° Inclined,
Interpupillary Distance 47-78 mm
WF 10X/20 and P16X/11
Backward Quadruple Nosepiece
Achromatic Objective/Infinite Semi-plan Objective 4×, 10×, 40×, 100×
Coaxial Coarse and Fine Adjustment,
Moving Range 0.002mm, Fine Division 20mm
Abbe Condenser, NA1.25
Rackless Double Layers Mechanical Stage 150×139mm,
Moving Range 75×52mm
1W S-LED
1
fig. 19
fig. 20
24
1
2
3