Herunterladen Inhalt Inhalt Diese Seite drucken

Battery & Product Disposal Information - miniland TH2071 Bedienungsanleitung

Stirn- und ohrthermometer
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 72
a Field strengths from fixed RF transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theore-
tically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromag-
netic site survey should be considered. If the measured field strength in the location in which device is used
exceeds the applicable RF compliance level above, device should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
device.
b
Over the frequency range 0.15 MHz to 80 MHz, field strengths should be less than 3 V/m.
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are contro-
lled. The customer or the user of device can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and device, as recommended
below, according to the maximum output power of the communications equipment.
Rated
maximum
output
power of
0,15 MHz to 80 MHz
transmitter
W
0,01
0,12
0,1
0,38
1
1,2
10
3,8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
8. BATTERY & PRODUCT DISPOSAL INFORMATION
• Remember to dispose of the product responsibly.
• Do not mix products bearing the crossed -out bin symbol with your
general household waste.
• For the correct collection and treatment of these products take them
to the collection points designated by your local authority. Alternatively
contact the retailer who sold you the product.
Recommended separation distances between
portable and mobile RF communications equipment and device
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
0,12
0,38
1,2
3,8
12
m
80 MHz to 2,7 GHz
37
0,23
0,73
2,3
7,3
23

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Thermotalk plus 89068

Inhaltsverzeichnis