6.3. WARNING OF DIRT ON THE SENSOR
In order to remind the user to clean the infrared sensor, this thermometer
has a special alarm. If the thermometer hasn´t been used for a long time
or if several measures have been taken, the LCD will display the cleaning
symbol to remind the user to clean the infrared sensor. The thermometer will
return back to standby mode after pressing any button.

6.4. CARE AND CLEANING

1. Keep the probe surface clean, otherwise the reading will not be accurate.
2. Clean the probe with clean paper moistened in water or alcohol, and take
the reading after the water on the surface of the probe all evaporates.
Caution:
Make sure the tip of thermometer is clean. A dirty lens may cause an
inaccurate reading.

7. TECHNICAL SPECIFICATION

• Product Name: Ear and forehead thermometer
• Model: TH2071 (thermotalk plus 89068)
• Power supply: DC3V (2 x AAA battery)
• Mode of operation: Adjusted mode
• Reference body site: Core
• Temperature sensing Part: IR sensor
• Range: 32.0°C ~ 42.9°C (89.6°F ~109.3°F)
• Room Temperature: 5.0°C ~ 59.9°C (41.0°F ~139.8°F)
• Laboratory accuracy:
+/- 0.2°C (or +/- 0.4°F)
35.5°C ~ 42.0°C (or 95.9°F ~107.6°F)
+/- 0.3°C (+/- 0.5°F)
32.0°C ~ 35.4°C (or 89.6°F ~ 95.8°F)
42.1°C ~ 42.9°C (or 107.7°F ~ 109.3°F)
• Room Temperature: +/- 2°C (4°F)
• Size: 127(L) x 39(W) x 63(D) mm
• Weight : 67 g
• Working environment: 10°C~40°C (50.0°F~104.0°F) for Ear
15°C~40°C (59.0°F ~104.0°F) for Forehead
Relative humidity: 20%RH~85%RH
• Transportation and storage condition:
Temperature -20ºC to 55ºC, Relative humidity 15%-90%
• Precision/repeatability: +/- 0.3°C (+/- 0.5°F)
• Service life of the battery: With a new battery (Carbon) approximately 6
months based on a frequency of use of 5 timer per day.
• Service life of the device: 2 years
*Dispose of used batteries responsibly, please consult your distribution center for
details.