Pierenkemper Operating Manual - PierenFemme

Warranty

Legal right of warranty is applied according to German Civil Code.

Guarantee

The manufacturer issues a guarantee of 12 months from the date of purchase.
The guarantee does not apply in the following cases:
»
for wearing parts and consumable supplies like, for instance, electrodes,
batteries and cables.
»
damage caused by the customer
»
defects the customer is aware of on the date of purchase
»
damage due to improper handling

Care and cleaning

No particular care and cleaning is necessary for the PierenFemme. The device and
cables can be cleaned with a soft, fluff-free cloth. It is important that no moisture
penetrates the interior of the device. If, nevertheless, moisture is present, the
device must be subjected to a technical check.

Classification

The above named product has been classified as Class IIa in accordance with
Appendix IX of the Directives 93/42/EEC for medical products.

Technical safety checks

We recommend a technical safety check on the PierenFemme every 60 months.
The technical safety check comprises:
1. Checking that the accompanying documentation is at hand and contains the
operating instructions together with the medical product book.
2. Checking the equipment for completeness.
3. Sight check
· For mechanical damage on the device and accessories,
· Completeness of stickers on the device,
· If the keys are working properly.
46