Checking the device functions
The device function must be checked before each use and after each re-use
/ preparation. Also note the summary of information on start-up on page 8.
Switch the device ON at the switch (1) on the rear for this purpose. Press
the ON/OFF button (10) on the control panel and start nebulization.
Check the mist quantity output using the keys for the mist quantity. The
highest quantity of mist should be released on setting 5.
Change the air flow using the keys for the air quantity. The highest air flow
should be released on setting 5.
Detach the transducer head from the device. The device goes into a
malfunction state, the red LED flashes and a warning signal sounds. Switch
off the device, replace the transducer head and switch the device on. The
nebulizer should resume its operation.
We recommend that the ultrasonic nebulizer undergoes a technical
safety inspection once a year by authorized personnel or the
manufacturer to ensure electrical safety.
 
 