Malfunction indicators
The display panel (8) for any error messages is on the right side of the
control panel (9). In the event of a malfunction, the device switches off
power and emits an audible warning. The nature of the malfunction is
indicated. If a malfunction is indicated, first switch off the device (push
ON/OFF button for approx. 1 s)
If possible, remedy the cause of malfunction and switch on the nebulizer
again.
Displays
+
+
+
+
+
+
Possible cause
Transducer head
incorrectly inserted
Contact liquid in
transducer has not been
refilled
Aerosol empty
Internal temperature too
high
page 13
Remedy
Mount transducer
head correctly.
Note the
anti-rotation
device
Fill contact water
up to blue mark or
replace nebulizing
chamber
Replace nebulizer
chamber
Turn device off
and allow it to cool
down