Intended use

Vestfrost biomedical freezers are designed to provide reliable temperature stability and precision
and ensure an ideal cold temperature storage for sensitive items is achieved� The products are
designed for the following operating ranges:
Models
Temperature range
Fabric Set Point
Ambient Temperature
Relative humidity
Number of probes
WARNING: This unit is not a "rapid-freeze" device� Freezing large quantities of liquid,
or high-water content items, will temporarily increase the chamber temperature and will cause
the compressors to operate for a prolonged time period� Avoid opening the door for extended
time periods since chamber temperature air will escape rapidly� Also, keep the inner doors closed
as much as possible� When room air, which is higher in humidity, replaces chamber air, frost may
develop in the chamber more rapidly�

Room climate classes

The climate class is stated on the nameplate� This specifies the optimum room temperature�
ULTF40/ VTS098
VTS252
-60°C to -86°C
-15°C to -30°C
-82°C
+16ºC to +32ºC
+16ºC to +32ºC
MAX 70 %
MAX 70 %
1
Room climate
class
3
4
5
7
VTS254
-25°C to -45°C
-25°C
-45°C
+16ºC to+ 32ºC
MAX 70 %
1
1
Max room
Relative
temperature
humidity
°C
25
30
40
35
Vestfrost Solutions
VTS256
-40°C to -60°C
-61°C
+16ºC to +32ºC
+16ºC to +32ºC
MAX 70 %
1
%
60
55
40
75
9
VTS258
-60°C to -86°C
-82°C
MAX 70 %
1