Caution!:
Please handel your scale with care.
Take notice of the following recommendations.
* Liquids (like water) which come in contact with the balance can harm the accuracy.
* Overload of the balance can harm the accuracy of the balance.
* Extreme temperature changes can harm the accuracy of the balance.
* A clash with a heavy machine can make the balance useless.
* Changes in the balance may only be done by your supplier.
* We recommend you to let the balance check by your supplier every year.
* Only trained employees are allowed to use the balance.
* If you warm up the balance for 15 minutes before using, the accuracy of the balance will be better.
* Only use the balance if it is locked rightly and has a serial number.
The balance can be sealed with a stamp or a sticker. Do not remove them from the balance.
* Always use the parts of your supplier.
Calibrate the scale:
When you have a standard weight (available from your supplier)
you can calibrate the scale yourself:
* Turn off the scale.
* Move the calibration switch at te rear from the indicator to position UNLOCK
* Switch On the scale, the display shows SPN
* Press TARE key, the scale-capacity appears in the display
* Put this Massweight on the scaleplatform and press TARE key to confirm
* After the BEEP the displays shows the correct weight, remove the Massweight
* Move the calibration switch at te rear from the indicator to position LOCK
* Install the scale on stable surface and switch on the scale.