Malfunction
Li-ion battery cannot be
loaded
Downloaded from
www.Manualslib.com
Detection
Li-ion battery in place,
charge state indicator
does not light up
manuals search engine
Cause
Remedy
Charger contacts dirty or
Clean contacts in the
damaged
charging bay, see Chap-
ter Cleaning/disinfection
or send to manufacturer
for repair
Li-ion battery defect
Send Li-ion battery to
manufacturer for repair
Charger defect
Send charger to manu-
facturer for repair
Excess
temperature
is
Remove Li-ion battery
detected in the Li-ion
from charging bay, leave
battery during charging
to cool and start the
charging process again.
If the problem occurs
again, send battery to
manufacturer for repair
Charging fault or
Remove Li-ion battery
Li-ion battery defect
from charging bay and
repeat charging process,
if necessary in another
charging bay.
If the problem occurs
again,
send Li-ion battery to
manufacturer for repair