9. Testing Procedure

Samples should be analyzed as soon as possible after collection. Exposure to
sunlight, other strong light, or agitation will accelerate the reduction of chlorine.
Use the measuring cup to fill the cell with 10 ml of sample liquid then screw the cap on.
Use a lint free soft cloth to dry the cell (7) and wipe any fingerprints off before inserting
into the cell chamber (4).
Press
to turn the meter on.
Press and hold
Put the cell (7) into the chamber (4), and
line the dot on the cap (6) with the triangle
on the meter.
Press and hold
adjust to zero.
After the reading (D) stops flashing and
shows "0.00", remove the cell (7).
Unscrew the cap (6) and insert one FCL
(DPD-1) strip or one TCL (DPD-4) strip
into the sample solution.
Fan the strips in a gentle back and forth
motion (approximately 2 strokes per
second) for 20 seconds, then remove the
strip.
Screw the cap (6) back on and wipe the cell (7) dry, taking care not to leave any fingerprints
on the glass.
Bubbles can lead to incorrect measurements. If needed gently swirl the cell (7) to
remove any bubbles before testing. Do not agitate the sample.
Insert the cell (7) into the chamber (4), lining the marks up as shown above.
Press
to get a reading.
Thoroughly rinse cells (7) and caps (6) with distilled water after testing. If reagent is
left inside for too long they may stain the glass which will affect accuracy.
24
for two seconds to select FCL (B) or TCL (C) mode.
for two seconds to