which will be displayed on the screen as
C. To avoid discomfort with the change
in programme, the set intensity will drop
by half. This programme has 60 steps
of intensity. Press and release the ▲
button until you reach a comfortable
intensity setting. This is a very natural
way for you to control your pain as you
rapidly increase and decrease the
intensity to match the strength of your
contractions.
Again, there are two modes, C and C
UP (Active Power Ramp mode).
However, this time mode C is constant.
Pressing the BOOST button starts the
Active Power Ramp mode which
ramps the power or level of intensity
during the final contractions to give the
additional "Brief intense" pain relief.
In between contractions use the ▲ and
▼ buttons until you reach a comfortable
level.
contraction just press the BOOST
button down and HOLD IT.
The intensity will increase until you
release the BOOST button because the
intensity may become too much to bear.
When you release, the intensity will
drop
background setting you pre-selected.
As it falls, you can press and hold the
BOOST button again to keep the
intensity at a fixed level to the end of
your contraction.
When instructed to start 'pushing' you
can use the programme button P to
switch perfect mamaTENS back to
programme A.
Downloaded from
www.Manualslib.com
When
you
experience
rapidly
and
return
manuals search engine
Just remember the three programmes
A, B and C which you can choose at
any time as your labour progresses,
and when you need a boost, just press
the BOOST button.
TENS can be used for as long as it is
necessary. Continuous treatment is
fine, but, except during labour, the
electrode pads should be repositioned
regularly (about every 12 hours) to
allow the skin to be exposed to the air.

8. ELECTRODE PADS

The electrode pads must always be
used in pairs (two electrode pads on
each channel), so that the signal can
flow in a circuit.
The electrode pads need to be placed
over the junctions of the nerves joining
a
the womb and birth canal to the spinal
cord. Correct positioning of the pads will
give maximum pain relief so please
study the images in section 11 and then
follow these instructions.
Always check that the unit is OFF
before
electrode pads.
Perfect mamaTENS resets strength
to zero if the pad or lead is
to
the
disconnected from your body. This
is to prevent sudden changes in
sensation when the pad is re-
connected.
ELECTRODE PADS ADVICE:
• The electrode pads supplied are
reusable but for single patient use.
12
attaching
or
removing