should be allowed to discharge until the battery low warning is
activated and then fully charged as in the following procedure.
a. Plug the adapter into the AC outlet and connect the cable to
the printer.
b. The Charging light at the front of the keypad will be
illuminated while the batteries are being charged.
c.
The battery pack will take 16 hours to become fully charged.
d. Warning! Use only the AC adapter supplied. Use of any other
type may cause permanent damage to the PrinterNOx
may cause a fire or electrical hazard.
Notes:
Do not leave the PrinterNOx on with discharged batteries for
long periods since this may result in deterioration of the
batteries.
The room temperature should be between 5 to 40 degrees C
during charging.
The printer can be used immediately after the AC adapter is
connected, however, if the battery pack is not fully charged,
printing may stop temporarily until sufficient charge is
available.
Do not short-circuit the battery pack.
Batteries at the end of their service life should be returned to
CareFusion or an authorised representative for correct
disposal. These must only be replaced with the approved part
(CareFusion part No. 36-BAT1033)
The PrinterNOx is CLASS I equipment (mains earth
connected).
Warning! Do not puncture or incinerate the rechargeable
batteries.