When the occlusion is removed, the display will automatically revert
to the measurement screen. Repeat this procedure by occluding the
exhaust port.
The sampling line should now be connected to the administered gas,
normally as close to the patient as possible. The sampling port must
be positioned in such a way that water cannot enter the tubing,
causing rapid filling of the water trap.
If the gas line appears to be continuously occluded, remove the
water trap and empty any water into a suitable container. Ensure
that there is no occlusion on the exhaust port.
Under exceptional circumstances, water may pass through a
damaged water trap and obstruct the inlet to the PrinterNOx. In this
case turn the unit off and draw any liquid out of the inlet with a
syringe connected directly to the inlet. Replace the water trap with
a new one, turn the unit on, and return to the measurement screen.
On the left-hand side of the NO legend are the upper and lower
alarm levels for NO. On the left-hand side of the NO
legend is the
2
upper alarm level for NO
.
2
On the bottom line, the rectangle will blink on or off every time a
measurement is updated (every 2.6 seconds) to indicate correct
functioning of the instrument.
The gas concentrations are given in parts per million (ppm).
If the printer option is turned on (see Setup section), then the
following printout will be obtained during measurement: