Screw the control valve on to the canister and connect to the
PrinterNOx with the water trap connected as shown below:
Pre-set
Flow
Regulator
(500ml/min)
Pressure
Guage
0
1000
psi
USE NO OIL
Control
Valve
25 ppm
Nitric Oxide
in Nitrogen
OR
10 ppm
Nitrogen Dioxide
in air
Slowly turn the control knob fully anti-clockwise. This will then
supply a flow of approximately 500ml/min. As the PrinterNOx
samples at a rate of 250ml/min then 250ml/min will be vented
through the one way valve.
One Way
Relief Valve
Micro
Medical
Alarm
20
PrinterNOx