Battery Cover
2X Locking
tabs
Caution
Replace the batteries in the transmitter with high quality "AA" size alkaline
batteries as shown. Never mix manufacturers when replacing the batteries.
Never mix old and new batteries. Care must be taken when replacing the
batteries not to damage the o-ring seal on the battery cover. The o-ring seal
should be replaced whenever it is damaged or its integrity is in question.
Arthrex Inc. recommends the seal should be replaced at the minimum every
third time batteries are replaced. When replacing the seal, lubricate it with
petroleum jelly.
Battery Leakage- If the transmitter or receiver will not be used for an
extended period of time, remove the batteries to prevent damage due to
possible battery leakage.
Battery disposal- Follow the battery manufacturer's recommendations or
your facility's policy for the disposal of used batteries.

Battery Replacement

Transmitter
P/N LM0659 Rev. 3
Slotted drive ¼
turn fastener
"O" Ring Seal
Recessed area for cover
removal. Gently pry with
flat tip device to remove
cover.
3X AA Batteries
3X Battery polarity indicator
on bottom face of holder