Hardware ReferenceÑFlextight Precision II
26
The Light Tubes
Figure 8:
To replace a tube:
1.
Get the appropriate replacement lamp (reßective or transmissionÑ
they differ only by the type of sleeve they have). Both lamps are
OSRAM L 8W/12-950 Lumilux De Luxe Daylight (5400K) lamps.
but each has a different customized reßective sleeve.
2.
Turn off and unplug the scanner.
3.
Remove both side panels:
¥ Get an allen key or some other long, thin, stiff instrument.
¥ Insert the instrument into the slot at one of the top side corners
¥ Place your hand against the side panel so that it will not fall
¥ Gently press the instrument against the side panel until it comes
¥ Repeat for the other side panel.
Reßective Lamp
Align the sleeve with the
pins as shown. The writing
on the light tube must be
under the sleeve as much
as possible.
Sleeve alignment. The relative position of the two lamps is
approximately as seen from the left side of the scanner.
of the recessed area at the back of the scanner (see Figure 2). It
will slide in until it touches the inside of the side panel.
when it comes loose.
loose from the magnets and falls into your hand.
Illuminate CCD
Transmission Lamp
Align the sleeve with the pins as shown.
Note that the long top slot should line up
with the two pins.
Imacon Flextight Precision II UserÕs Guide
Illuminate
light table