General description
The Sylvac external micrometer S_Mike is equipped by a sensor SIS (Smart Inductive
Sensor) with maintenance of reference position, even when the instrument is in standby
mode. Its principle with non-rotary spindle allows a very fast positioning.
It includes the following functional features:
Input of Preset values.
Selection of the measuring unit (mm / INCH).
Transfer of measuring value (Proximity connection).
Protection against penetration of dust and splashing of liquids (oil, water), protection
degree IP67 according to (IEC 60529 standard). This protection degree is also
guaranteed when using the Proximity connector.
The micrometer S_Mike is extremely easy to operate because of its menu selection system.
Switching ON, Initialization
During the first start up, the micrometer displays the measuring unit selection mode mm
(metric) or IN (inch). Press the key below the required measuring unit to set it as the default
one.
After having introduced the measuring unit, the instrument shows "SET" for the initialization
of the reference point. Bring back the measuring spindle on the anvil (or on a Gauge block).
Press then [SET] or [MODE]. The instrument is now ready (Check the Preset value, if
necessary change this value and reinitialize the instrument, see "Preset mode").
Standby
Mode of reduced consumption without loss the reference position. The standby mode is
activated automatically after 20 minutes of no use. It can also be activated by pressing the
[SET] key until extinction of display.
The instrument awakes automatically in case of movement detection on thimble, by a
pressure on a button or when requesting Data.
Full switch off
In order to fully switch off the instrument (minimum power consumption), press the [SET]
button until the message "OFF" appears. At this state no data is maintained and therefore
the Reference position is lost. The "SET" message will be displayed again during the next
use (see "Switching ON").
Description of the menu system
The [MODE] key enables the selection of the different menus (each key stroke activates the
following menu). The [SET] key enables the activation of a function assigned to the
corresponding selected menu.
To avoid a wrong action, each menu is active only during 5s. This period passed, the
instrument switches automatically back to "Measuring Mode" as long as no action has been
performed to the [MODE] or [SET] key.
Preset mode
Preset makes it possible to introduce a reference value different from zero.
Press many times the [MODE] button until menu "PRE" appears. Then press the [SET]
button to activate the function Introduction of Preset. The display shows 00.000 or the last
stored value of Preset. A cursor is displayed under the first digit.
Each push on the [SET] button modifies the value of the digit above the cursor (±, 0..9).
Each push on the [MODE] button moves the cursor to the right. A long press on the [MODE]
button validates the value of Preset and leaves the Preset menu. The instrument
automatically returns to "Measuring Mode" when there are no further actions regarding the
buttons [MODE] or [SET] for a period of 15s.
Changing the measuring unit (mm/IN)
Press many times the [MODE] key until the menu "UNIT" is displayed. Then press the [SET]
key to change unit: "MM" or "INCH".
Locking of the instrument
Press [MODE] until the display shows " LOC". Then press [SET] to lock the instrument.
If the instrument is locked, the function send data (button [SET]) only is active. A pressure of
5s on a button, unlock the instrument.
Transmission of Data
The external micrometer has a contact less data output. This data output enables the direct
connection to a computer or a printer using a Proximity connection cable.
A short press on the [SET] button transmits the current value to the data connection. The
indicator send Data is briefly displayed. For more information, see the instructions of the
Proximity cable.
Reset, recall of the Preset
A push of 2s on the [SET] button assigns the Preset-value (or zero) to the current position of
the measuring spindle.
Remote commands
The data request is done by the sending of the character ASCII <?>, followed by <CR>
(carriage return). The instrument also answers in the same way during the detection of an
impulse pedals/automat.
The other commands are shown in the following table:
BAT?..............................Sends the status of the battery (BAT1=ok or BAT0=low)
FAC RST .......................General Reset (restores the factory parameters, NUM=276)
IDE or ID ?.....................Sends the instrument identification (SY276)
IN ...................................Activates the Inch unit
KEY? .............................Sends keyboard status (KEY0 or KEY1)
KEY 0 or OFF ................Locks the keyboard (except data output)
KEY 1 or ON..................Unlocks the keyboard
LCAL?............................Sends the date of the last calibration (DD.MM.YYYY)
LCAL DD.MM.[YY]YY ....Sets the date of the last calibration
MM.................................Activates the millimeter unit
NCAL? ...........................Sends the date of the next calibration (DD.MM.YYYY)
NCAL DD.MM.[YY]YY ...Sets the date of the next calibration
NUM ? ...........................Sends the instrument number
NUM XXXX....................Modifies the instrument number (0...9999)
ON .................................Wake-up of the instrument
OFF ...............................Standby mode of the instrument
OUT 0 or OFF................Disables the automatic data transmission
OUT 1 or ON .................Enables the automatic data transmission
PRE? .............................Sends the Preset value (±xxx.yyy[yy])
PRE ±XXX.YYYYYY......Sets the Preset value (max. 200.0mm / 8.00'')
P or PRI .........................Sends the displayed value
RST ...............................Reset of the instrument (user's parameters)
SET?..............................Transmits the configuration of the instrument (IN or MM,
.......................................STO0 or STO1, KEY0 or KEY1, BAT1 or BAT0)
STO? .............................Sends the display status
STO 0 or OFF................Unlocks the display
STO 1 or ON..................Locks the display
UNI? ..............................Send the current unit of measure (IN or MM)
UNI 0 or OFF .................Locks the Unit function
UNI 1 or ON...................Unlocks the Unit function
VER? .............................Sends the version and date of software (Vx.x DD.MM.YYYY)
Error messages
In case of problems, the following error messages will be sent on the S_Proximity
connection and on the display :
- ERR0 : Error of the reading sensor
- ERR1 : Error of parity
- ERR2 : Error in the received remote command
- ERR3 : Overflow of the measuring range
- ERR8 : Memory error
- ERR9 : Critical error, the instrument must be revised
The errors 0 and 3 must be confirmed by the button [MODE] or [SET], and then the
instrument must be reinitialized.
Replacing the battery
The display of the letter "B" indicates the end of the battery life. However there remain still
some working hours.
Maintenance
Keep the micrometer in a dry environment when not using it for a longer period of time to
avoid rust formation of the metallic parts.
Do not close the measuring spindle with the anvil when not in use. Keep a distance of 1-2
mm.
Do not use aggressive products (alcohol, trichloroethylene or others) to clean the plastic
parts. Do not keep the micrometer in places which are exposed to sun, heat or humidity.
Important : dry carefully all metal parts of the instrument after effect of moisture to
guarantee a perfect mechanical functioning and to avoid rust formation.
Technical specifications
Measuring range .................................. 0 - 30mm | 30 - 66mm | 66 - 102mm
......................................................... 0 - 1.2''
Resolution ............................................ 1µm | .00005"
Accuracy .............................................. 0 - 30mm :
......................................................... 30 - 66mm : 4µm | .00016"
......................................................... 66 - 102mm : 5µm | .0002"
Measuring force ................................... 0 - 30mm : 10N, adjustable to 5N
......................................................... 30 - 66mm : 10N ( 5N on request)
......................................................... 66 - 102mm : 10N ( 5N on request)
Repeatability ........................................ 2µm | .0001"
Flatness of measuring faces ................ 0.6µm | .000025"
Parallelism of measuring faces............. 0 - 30mm :
......................................................... 30 - 66mm : 2µm | .0001"
......................................................... 66 - 102mm : 3µm | .00012"
Spindle travel per thimble revolution ... 12mm
Number of display sequences ............. 8 /s
Measuring system ............................... Sylvac inductive system (patented)
Power supply ....................................... 1 lithium battery, type CR2032
Mean power consumption ................... 50µA (continuous)
Stand-by power consumption .............. 8µA
Mean battery life (automatic standby
with saved origin) ................................. 12'000 hours
Data output .......................................... S_Connect Proximity
Data transfer parameters ..................... 4800 bauds, 7 bits, even parity, 2 stop bits, no flow-
......................................................... control
Operational temperature ...................... +5° to + 40°C | +41°F to +104°F
Degree of protection ............................ IP67 (IEC 60529)
Weight ................................................. 270g (0 - 30mm) | approx. 425g (30 - 66mm) |
......................................................... approx. 550g (66 - 102mm)
Other specifications .............................. according to DIN 863-1
Code number
Instrument 0 - 30mm ......................................................................................... 903.0300
Instrument 30 - 66mm ....................................................................................... 903.0600
Instrument 66 - 102mm ..................................................................................... 903.1000
Set of 3 instruments 0 - 102mm ....................................................................... 903.1300
Cable S_Connect Proximity-USB ..................................................................... 926.6721
Cable S_Connect Proximity-RS232 .................................................................. 926.6723
Stand ................................................................................................................. 903.0330
50mm Gauge Block for instrument 30-66mm..................................................... 903.0620
75mm Gauge Block for instrument 66-102mm................................................... 903.1020
E
| 1.2 - 2.6''
| 2.6 - 4''
3µm | .00012"
2µm | .0001"