Key-operated switch (for centrifuges with key-operated switch only).

The following program interlocks can be set by the key-operated switch:
Left key position:
Right key position:
Middle key position:
27

Program linking (for centrifuges with program linking only).

Several centrifugation runs can be linked together with the help of the program linking function.
27.1

Linking programs

Program linking is only possible with programs in which run-up and braking steps have been set.
Before linking, the programs must be stored in the desired sequence either by entering the program or by
calling up the program (see Chapter "Programming").
The program locations must be consecutive (e.g. program locations 10+11+12).
1.
Press the
PROG
2.
Set the program location for the first program (XX+) by means of the control knob
3.
Press the
RCL
4.
Press the
PROG
5.
Press the
STO
(+XX+) will be displayed.
6.
Press the
RCL
7.
Press the
STO
(+XX+) will be displayed.
8.
Keep repeating steps 6 and 7 until all programs have been linked.
9.
Press the
PROG
27.2

Changing program links

Call up the desired program (see Chapter "Calling up programs"), change the desired parameters (see Chapter
"Entering centrifugation parameters"), and save the modified centrifugation data to the same program location
(see Chapter "Entering / modifying programs"). As a result of saving, the program linking will be removed.
Re-link the programs (see Chapter "Linking programs").
27.3

Centrifugation run with program linking

Press the
PROG
Set the program location for the first program (XX+) by means of the control knob
Press the
RCL
Press the
START
The run-up and braking steps for the program linking will be displayed:
x
PR-PART
x
PR-PART
y
x
PR-PART
When the time in the last program has expired, run-down will take place with the braking step of the last
program. If the centrifugation run is interrupted by pressing the
braking step of the program running at the time.
40/102
LOCK 1 will be displayed.
Programs can be called up but not changed.
LOCK 2 will be displayed.
No programs can be called up or changed.
no status display.
No program interlock. Programs can be called up and changed.
key in order to select the parameter PROG-Nr. The input field will be displayed inversely.
key. The centrifugation data for the selected program location will be displayed.
key twice in order to select the parameter PR-PART. The input field will be displayed inversely.
key twice. The program will be linked and the program number for the next program location
key twice. The centrifugation data for the selected program location will be displayed.
key twice. The program will be linked and the program number for the next program location
key to exit. The program number for the last program (+XX) will be displayed.
key twice in order to select the parameter PR-PART. The input field will be displayed inversely.
key. The centrifugation data for the selected program location will be displayed.
key. The rotation indicator
First program (XX+)
x: Run-up step for the first program
Next program (+XX+)
x: Run-up step for the next program
Last program (+XX)
x: Run-up step for the last program
y: Braking step for the last program
will appear as long as the rotor is turning.