Troubleshooting - Cresta BPM 220 Bedienungsanleitung

Automatische handgelenk blutdruckmesser
Vorschau ausblenden Andere Handbücher für BPM 220:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Measuring Range
Pressure: 30~280 mmHg
Pulse: 40~199 Beat/min
Accuracy
Static Pressure: ± 3mmHg Pulse: ± 5%
Memory
90 Memories
Power supply
2x 1.5V Batteries (LR03 or AAA)
use alkaline battery, measure above 200 times
Operating condition:
+5
Atmospheric pressure: 70kPa~106kPa
Storage condition:
-20
Atmospheric pressure:50kPa~106kPa
Dimensions
Approx.: 62(W)X78(H)X31(D)mm
Weight:
Approx.: 130g, excluding batteries
Type BF
Classification
13,5 ~19,5 cm
Wrist cuff
Specifications may be changed without notice in the event of improvement
being made.
1.Type of protection against electric shock: INTERNALLY POWERED EQUIPMENT.
2.Degree or protection against electric shock: TYPE BF APPLIED PART.
3.Mode of operation: CONTINUOUS OPERATION.
4.Equipment not suitable for category AP&APG equipment use in presence.
STATEMENT
The system might not meet its performance specifications if stored or used
outside the temperature and humidity as mentioned below:
Operating conditions: +5℃ ~+40℃ 15%RH~93%RH 70kPa~106kPa
Storage conditions: -20℃ ~+55℃ . 0%RH~93%RH

TROUBLESHOOTING

lf you have trouble in using the unit please check the following points first.
ERROR DISPLAY
POSSIBLE CAUSE
No battery installation
Nothing is displayed
when you push the
Battery worn out
POWER button or the
The polarities of batteries
battery icon flashes
placed wrongly
E1: can't normally
Check your cuff if any
Increase pressure
air leakage
E3: inflate pressure too
high
℃ ~+ ℃
40 .
15%RH~93%RH
~+55 . 0%RH~93%RH
.
HOW TO CORRECT
Insert batteries
Replace new batteries
Insert battery in the
correct polarities
Replace the cuff with
new one
Re-measure or send
back to the dealer for
re-calibrate pressure
12
E2E4: have shaking
Hand or body shaking
while measurement
while measurement
Battery icon on
Battery low power
The systolic pressure
1.The arm cuff was held
Value or diastolic
lower than your heart
Pressure value too
2.The arm cuff was not
high
attached properly
3.You moved your body or
spoke during measurement
1.The arm cuff was held
The systolic pressure
higher than your heart
Value or diastolic
2.you moved your body or
Pressure value too low
Spoke during measurement
Appendix 1 Guidance and Manufacturer Declaration Tables
Guidance and manufacturer's declaration – electromagnetic emissions
The Model BPM220 Series Electronic Blood Pressure Monitor is intended for
use in the electromagnetic environment specified below. The customer
or the user of the Model BPM220 Series Electronic Blood Pressure Monitor
should assure that it is used in such an environment.
Compliance Electromagnetic environment-guidance
Emissions
The
RF emissions
Group 1
Pressure Monitor uses RF energy only for its
CISPR 11
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
Class B
CISPR 11
Harmonic
emissions
N.A.
The
IEC 61000-3-2
Pressure Monitor is used in home and it's
powered by DC 3V
Voltage
fluctuations/
flicker
N.A.
emissions
IEC 61000-3-3
keeping static and correct
gesture to measure again
Replace battery and
measure again
keeping correct
position and gesture
to measure again
Model BPM220
Series Electronic Blood
Model BPM220
Series Electronic Blood
13