WARNING SCENARIO TABLE:

SCENARIO
Start up
Standard running
Low battery level
Discharged battery
Blocked filters
Other fault
DOFFING: (DO NOT remove the product whilst in the hazard zone)
1. Pull the tabs on the seal-skirt outwards and upwards to raise the visor.
2. Turn the red adjustment wheel anti-clockwise to loosen the helmet. Lift the assembly off the head.
3. Press and hold the power button on the battery pack to switch off. The adjacent battery charge level indicator will light momentarily to
display the remaining charge level.
1

MAINTENANCE:

It is important to follow these instructions to ensure that the PowerCap
If the maintenance and usage instructions in this leaflet are not followed, the effectiveness and the protection offered by the PowerCap
may be severely reduced. A Radio Frequency Identification Device (RFID) has been built in to the product to allow maintenance records to be
kept, reviewed and amended by your own product support representatives using JSP's Asset Management system. This allows you to have
full control over your PowerCap
much more. Visit www.jspcheck.com to enable your RFID device. To view a playlist of instructional videos visit bit.ly/Infinity-Videos.

FILTERS:

• The filters provided are high efficiency particulate types supplied in pairs.
• Will filter spores, fibres, fumes (very fine metallic droplets/particles) and dusts.
• Will not filter gases or vapours.
• PowerCap
Infinity
filters are marked as per image.
®
®
Only use genuine PowerCap
®
JSP Ltd or its distributors. Always replace both filters simultaneously.
NEVER use compressed air to clean the filters.
This will seriously limit protection.
DISPLAY
Red, green, blue
Green static
Green (Red flash every 2s)
Red flash
Blue flash
Red, blue (repeat)
2
TAB
Infinity
giving full access in the field to inspection dates, compliance documentation, product instructions and
®
®
Infinity
filters supplied in pairs from
®
BUZZER
On (2s)
Off
Off
High-low tone
High pulsed
High pulsed
Infinity
gives optimal performance. Regular maintenance is important.
®
®
9
COMMENTS
System OK
System OK
20 minutes remaining
Recharge battery
Change BOTH filters
Contact JSP Ltd
3
CHARGE LEVEL
POWER
Infinity
®
®

Werbung