Herunterladen Inhalt Inhalt Diese Seite drucken

Important Information; Ce-Label; Disposal - KaWe LED 2000 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für LED 2000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Temperature
Relative humidity
Air pressure
Instructions on the packaging
Temperature range
Humidity during
during transport
transport and
and storage
storage
+50° C
20% - 90%
–10° C

7.6 Important information

If multiple KaWe examination lights are used at once, the total
irradiance may exceed 1,000 W/m² due to superposition of the lumi-
nous fields. As a result, there is a risk that the illuminated field may
become very hot.
The superposition of the luminous fields of multiple KaWe examina-
tion lights may cause the limit values for UV radiation (<400 nm) of
10 W/m² to be exceeded.
The report on the factory test for electrical safety can
be obtained upon request. The only requirement is the serial
number of the light for which the protocol is desired.
If other KaWe examination lights or other pieces of equip-
ment are connected together during installation, section 16
of EN 60601-1:2013 applies and if necessary, the compliance
with the requirements is to be verified.
Before initial operation, the installation must be inspected
in accordance with EN 62353
Transport/storage
Min.
Max.
–10° C
+50° C
20 %
90 %
700 hPa
1060 hPa
Air pressure during
transport and
storage
RH
P
700hPa - 1060hPa

8. CE-LABEL

This KaWe examination light complies with guidelines
93/42/EWG (guidelines for medical products from the council of the
European communities). The applicable standard is EN 60601-2-41.
The company KIRCHNER & WILHELM GmbH + Co. KG is certified ac-
cording to DIN EN ISO 13485:2012.

9. DISPOSAL

At the end of the product's life, the components of the KaWe
examination light are to be disposed of properly. Ensure that the mate-
rials are sorted carefully according to type. The electrical circuit boards
are to be disposed of at an appropriate recycling centre. The lamp
housing and the rest of the components of the KaWe examination light
should be disposed of according to their type of material.
17