Herunterladen Inhalt Inhalt Diese Seite drucken

HACH LANGE QuikChem 8500 2 Serie Wartung Und Fehlerbehebung Seite 15

Vorschau ausblenden Andere Handbücher für QuikChem 8500 2 Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Problem
Areas and heights
The system has not
are shown, but there
been calibrated.
is no concentration
for samples.
Patterned air spikes
Time to valve is not
correct
Random air spikes
Dissolved gases
come out of the
solution
Transmission tubing
seal to nipple is not
complete
Bad reagent
Sampler initialization
Sampler fails to
error
initialize
The computer shows
More than one
that an application
program operates at
error or a runtime
the same time.
error has occurred.
The standards fail
Multiple causes
Calibration failure
Peak integration is
not correct
Standard preparation
is not correct
Placement of sampler
is not correct
Flow/chemistry
problem
Sampler time out
Problem with sampler
arm movement
Communication error
with the system
The heater element
Heater temperature is
does not get warm.
not set correctly.
Thermocouple is
loose or not
connected
Controller does not
work
Possible cause
Do a calibration and analyze the samples again.
If the air spikes occur with every peak in the same location, look
at the time for the sample to get to the valve and the load time.
• Prepare reagents a day in advance and keep the reagents at
room temperature.
• Degas all reagents with helium (unless otherwise noted).
• Use a longer backpressure loop. Increase in 50 cm
increments.
• See if the air is pulled in the reagent flow intermittently.
• Make sure that there is sufficient reagent volumes.
• Under the "Configuration" tab in the software, select
"Autosampler". Select "Initialize Autosampler".
• Make sure that the power to the sampler is on.
• Make sure that sampler cable is not loose.
• Make sure that the latest software is installed.
Select Ignore when asked. Close Windows
application again.
• Make sure that the check standards are made and added
correctly.
• Make sure that the peaks are integrated correctly.
• Make sure that the reagent preparation and placement are
correct.
• Make sure that the time to the valve is correct.
• Make sure that the sampler arm moves freely.
• Under the "Configuration" tab in the software, select
"Autosampler". Select "Initialize Autosampler".
• Look for loose or unplugged sampler cable.
• Make sure that the heater temperature has been set for that
channel in Run Properties>Analytes. Make sure that the set
point shows correctly in the Omnion channel data display.
• Connect the thermocouple
• Contact technical support
Solution
and start the
English 15

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis