CELSIUS (Med) / ESPRIMO (Med)
The EUT is intended for use in the electromagnetic environment specified below. The customer or
the user of the EUT Name should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Fujitsu
Table 2 – continued
Compliance
Electromagnetic environment - guidance
level
Portable an mobile RF communications
equipment should be used no closer to any
part of the EUT including cables, than the
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d
[3] V
d
[3] V/m
d
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, a should be less than the
compliance level in each frequency range. b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
3.5
=
P
V
1
3.5
=
P
E
1
80 MHz to 800 MHz
7
=
P
E
1
800 MHz to 2.5 GHz
English - 5