2. CCS–200 Shut Down
NOTE: For additional help on shutting down the system, please refer to the
frigitronics CCS-200 Operation and Maintenance Manual.
1. Turn the Nitrous Oxide Cylinder Valves clockwise to the "OFF"
position.
Turn the system On/Off Valves "ON" and "OFF" repeatedly until all
pressure gauges read "0" psi and the Exhaust Pressure Gauge
needles touch the silver stop.
2. Turn system On/Off Knobs to the "OFF" position.
CAUTION: Do not remove or install probes unless the line and exhaust
pressure gauge read "0" psi.
3. If attached, remove probe(s).
4. Turn Main Meter switch to the "OFF" position.
5. The CCS–200 shut down is complete.
3. Product Description
Important Preparation Notes:
It is recommended that the Tank Heater be installed and activated
ONE (1) hour prior to beginning of the surgical procedure.
Initial tank pressure range must be 925 - 950 psi. DO NOT allow tank
If the pressure exceeds 950 psi, unplug the Tank Heater from the outlet
If tank pressure drops below 925 psi prior to the beginning of the procedure,
plug the Tank Heater back into the power outlet until the starting pressure
This device is intended for use only with the AtriCure Frigitronics CCS200
1. The Tank Heater is temperature controlled (pre-set to 34ºC).
2. Each Tank Heater consists of three (3) components to be
assembled. See Figure 1.
approximately
pressure to exceed 950 psi.
reaches 925- 950 psi.
Console.