electromagnetic site survey should be considered. If the
measured field strength in the location in which the CPI-3
is used exceeds the applicable RF compliance level above,
the CPI-3 should be observed to verify normal operation.
c. Over the frequency range 150 kHz to 80 MHz, field
strengths should be less than 3 V/m.
Separation distances between RF
communications equipment and CPI-3
The CPI-3 is intended for use in an electromagnetic
environment in which radiated RF disturbances are
controlled. The customer or the user of the CPI-3 can
help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF
communications equipment (transmitters) and the CPI-3
as recommended below, according to the maximum
output power of the communications equipment.
Rated
maximum
output power of
transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not
listed above, the recommended separation distance d in
meters (m) can be estimated using the equation
16
Separation distance according to
frequency of transmitter (m)
d = 1.2√P < 800
MHz
0.12
0.38
1.2
3.8
12
d = 2.3√P ≥ 800
MHz
0.23
0.73
2.3
7.3
23