Electromagnetic immunity

The CPI-3 is intended for use in the electromagnetic
environment specified below. The customer or the user
of the CPI-3 should assure that it is used in such an
environment.
Immunity
Electrostatic
discharge
(ESD)
IEC
a
61000-4-2
Power
frequency
(50/60 Hz)
Magnetic
Field
IEC 61000-
4-8
14
IEC
Compliance
60601
level
test level
± 6 kV
± 6 kV
contact ±
contact
8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic
environment -
guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30 %. As with the
operation of other
electronic devices,
precaution should
be taken to not
generate ESD.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in
a typical
commercial or
hospital
environment.