1. Notice when using equipment

1. Keep the scaler clean before and after operation.
2. The handpiece, scaling tip, torque wrench, endo wrench and endochuck must be sterilized before each treatment.
3. Don't screw or unscrew the scaling tip and endochuck when stepping on the foot switch.
4. The scaling tip must be fastened and there must be fine spray or drip coming out from the tip when you use it.
5. Change to a new one when the tip and ultrasonic file become damaged or worn excessively.
6. While the scaler is working, the heat of scaling tip will become higher if there is no water flowing out. Please keep the water
flowing smoothly.
7. Don't twist or rub the tip and endochuck.
8. Ensure the connection of the handpiece and the socket of the cable are dry before installing the handpiece.
9. Don't pull the cable forcibly in case of the handpiece falling off from the cable.
10. Don't knock or rub the handpiece.
11. Ensure that the power supply can be easily removed in an emergency.
12. The power supply is considered as a part of the equipment. This device can only be equipped with the special
power supply supplied.
13. The power supply is NOT waterproof. Please keep it dry and away from the water.
14. After operation, turn off power, and pull out the plug.
15. We are only responsible for the safety on the following conditions:
- The maintenance, repair and modification are made by the manufacturer or the authorized dealer.
- The changed components are original of "CFPM" and are operated according to instruction manual.
16. The internal screw thread of the scaling tips produced by some manufacturers may be coarse, rusty or collapsed.
This will damage the external screw threat of the handpiece irretrievably.

2. Contra-indications

This device should not be used in patients with hemophilia.
Patients or practitioners who wear pacemakers should not use this device.
Caution should be exercised when this unit is used on children, pregnant women or people with heart disease.

3. Storage and maintenance

The appliance must be handled carefully, away from vibration and installed in a dry and ventilated place.
Don't store the machine together with the articles that are combustible, poisonous, caustic, or explosive.
This equipment should be stored in a room where the relative humidity is less than 80%, atmospheric pressure is 50 kPa to 106
kPa, and the temperature is -10°C to +50°C.
If not use for a long time, operate the unit with water at least once a month for 5 minutes.

4. Transportation

1. Excessive impact and shake must be prevented in transportation, lay it carefully and lightly and don't invert it.
2. Don't put it together with dangerous goods during transportation.
3. Avoid the sun or getting wet in rain or snow during transportation.

5. Working condition

Environment temperature:
Relative humidity:
Atmosphere pressure:
We offer one year warranty to the equipment according to the warranty card. The repair of the equipment should be carried out
by our professional technician. We are not responsible for any irretrievable damage caused by a non-certified technician.
GACD_280514_Livret TriscalerAqua_V17Bis.indd 9

5. PRECAUTIONS

6. AFTER SERVICE

9
5°C ~ 40°C
≤ 80%
70 kPa ~ 106 kPa
en