Table 1 – Guidance and manufacturer's declaration – electromagnetic emissions – for all ME equipment
and ME systems
ACP system is intended for use in the electromagnetic environment specified below. The customer or the user of ACP system should
assure that it is used in such an electromagnetic environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Table 2 – Guidance and manufacturer's declaration – electromagnetic immunity – for all ME
equipment and ME systems
ACP system is intended for use in the electromagnetic environment specified below. The customer or the user of ACP system should
assure that it is used in such an electromagnetic environment.
Immunity
test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60
Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level
Guidance and manufacturer's declaration – electromagnetic emissions
Compliance
Group 1
Class A
Class A
Complies
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601
test level
±8 kV contact
±15 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 s
3 A/m
Electromagnetic environment – guidance
ACP system uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The emissions characteristics of this equipment make it suitable for
use in industrial areas and hospitals (CISPR 11 class A). If it is used
in a residential environment (for which CISPR 11 class B is normally
required) this equipment might not offer adequate protection to
radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
Compliance
level
±8 kV contact
±15 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV mode differential
±2 kV mode common
0 Vac during 10 ms
92 Vac during 100 ms
161 Vac during 1 s
0 Vac during 5 s
3 A/m for 50Hz and for
60Hz
Electromagnetic environment
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Main power quality should be that
of a typical commercial or hospital
environment.
Main power quality should be that
of a typical commercial or hospital
environment.
Main power quality should be that
of a typical commercial or hospital
environment. If the user of ACP
system requires continued operation
during power mains interruptions, it
is recommended that ACP system be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial
or hospital environment.