gefran 2308 Bedienungsanleitung Seite 13

Mehrkanal
Vorschau ausblenden Andere Handbücher für 2308:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 18
Display
Press CFG for
CFG
10 sec.
Software
protection
Prot Configuration
Alarm
setting
0
Enabled
Enabled
1
Enabled
Disabled
2
Disabled
Enabled
3
Disabled
Disabled
if
= 0
Input
configuration
From 1 to last channel
Selection of
selected
channel
0= return to alarm
configuration phase
Selection of input type
0
J
0...800°C / 32...1472°F
1
K
0...1300°C / 32...2372°F
2
R
0...1600°C / 32...2912°F
3
S
0...1760°C / 32...2912°F
4
T
-100...400°C / -148...752°F
5
0...20mA
6
4...20mA
7
100m Vac (only HW config.)
8
0...20mV
9
0...50mV
10
0...100mV
11
0...10V
12 Pt100/2
-200...400°C / -328...752°F
13 Pt100/2
-99,9...100,0°C / -99,9...212,0°F
14 Pt100/3
-200...400°C / -328...752°F
15 Pt100/3
-99,9...100,0°C / -99,9...212,0°F
16
Potentiometer
Setting minimum
LINEARS
scale value
-999...9999 Scale range
Setting maximum
LINEARS
scale value
-999...9999 Scale range
filter 1,2Hz
0= filter ON
Digital filter
1= filter OFF
LINEARS
Setting
Press the "Raise" and
decimal point
"Lower" keys to move the
decimal point
Fixed decimal
Move from one configuration phase to another by pressing the CFG key; press the F key to go to display.
CAL
Display
If jumper S17 is
present
(see para. 7)
ATTENTION: the instrument is supplied calibrated; any additional calibration must be made by a trained technician equipped with the necessary
devices. The procedure is irreversible.
8c • Configuration
Serial communication
transmission speed
b
Baud rate
Protocol
Parity
0 1200 bit/sec
Cencal
1 2400 bit/sec
Cencal
2 4800 bit/sec
Cencal
3 9600 bit/sec
Cencal
4 1200 bit/sec Modbus
none
5 2400 bit/sec Modbus
none
6 4800 bit/sec Modbus
none
7 9600 bit/sec Modbus
none
8 1200 bit/sec Modbus
even
9 2400 bit/sec Modbus
even
10 4800 bit/sec Modbus
even
11 9600 bit/sec Modbus
even
12 1200 bit/sec Modbus
odd
13 2400 bit/sec Modbus
odd
14 4800 bit/sec Modbus
odd
15 9600 bit/sec Modbus
odd
if
configuration
C = 31 or 32
(AND/OR function)
alarm
No. first
Fixed
alarm
channel
No. last
alarm
TC/RTD
Selection of event
type for alarm
AND (C = 31)
Alarm active
0
if all alarms are
TC/RTD
active (ON)
Alarm active if at
1
least 1 alarm is
deactivated (OFF)
OR (C = 32)
Alarm active
0
if at least 1
alarm is
active (ON)
Alarm active
1
if all alarms
are deactivated
(OFF)
8d • Calibration
Selection of
Display type of input to be
channel to be
calibrated (1...8)
Instrument code
for digital
communication
Interface
Select a no. in range
CL/485
0...9999
CL/485
∆ ∇
485
with keys "
485
485
485
485
485
485
485
485
485
485
485
485
485
= 1...10
Alarm
Setting (alarm 1 and 2) and
display (alarm 3...10) of
channel assigned to alarm
C = 1...30
(normal alarm)
3...10
Setting value of
1...8
alarm limit
Absolute = scale range
Deviation = -999...9999
Setting alarm
t > F
hysteresis
value
(*)
Hyst 1 ÷ 2 = -999...9999
Hyst 3 ÷ 10 = -128...127
Setting alarm
event type
Absolute alarms
0
Alarm active
above limit
1
Alarm active
below limit
Deviation alarms
0
Alarm active above
(limit + absolute)
1
Alarm active below
(limit + absolute)
Symmetrical alarms
0
Alarm active
outside interval
1
Alarm active
inside interval
calibrated
(see para. 8c)
Display room
temperature setting *
for TC only
13
No. of channels
actually used
Set no. of channels
(1...8) actually used
(+8 to enable
automatic scanning)
"
Access to
configuration
phase
Set 0 to go to channel
configuration;
set 1...10 to go to alarm
configuration
All alarms
0
Alarm not engaged
1...8
Alarm assigned to channel 1...8;
alarm setpoint is absolute
Alarms 1 and 2 only
9
Alarm assigned to all channels; alarm
setpoint is absolute; alarm is active if all
channels are in alarm state
10
Alarm assigned to all channels; alarm
setpoint is absolute; alarm is active if at
least one channel is in alarm state
11...18 Deviation alarm assigned to channel 1...8;
alarm setpoint relates to setpoint of alarm
(3...10) assigned to channel
19
Alarm assigned to all channels; for each
channel, alarm setpoint relates to setpoint of
alarm (3...10) assigned to channel; alarm is
active if all channels are in alarm state
20
Alarm assigned to all channels; for each
channel, alarm setpoint relates to setpoint of
alarm (3...10) assigned to channel; alarm is
active if at least one channel is in alarm state
21...28 Alarm assigned to channel 1...8; alarm
setpoint defines a symmetrical event
window around the setpoint of alarm (3...10)
assigned to channel
29
Alarm assigned to all channels; for each
channel, alarm setpoint defines a
symmetrical event window around the
setpoint of alarm (3...10) assigned to
channel; alarm is active if all channels are in
alarm state
30
Alarm assigned to all channels; for each
channel, alarm setpoint defines a
symmetrical event window around the
setpoint of alarm (3...10) assigned to
channel; alarm is active if at least one
channel is in alarm state
31
AND function for multiple alarms (3...10)
32
OR function for multiple alarms (3...10)
Note:
if U=1, alarm tripping under channel
conditions OR and AND is reversed from
9 to 30.
(*) For symmetrical alarms with negative hysteresis,
hysteresis is placed inside the window, whereas with
positive hysteresis it is placed outside.
The minimum alarm (positive hysteresis, type 1 event)
is automatically excluded in the startup phase until the
input variable has exceeded the set limit.
No. of selected
channel
Transmit a signal equal
With a calibrator or
to max. scale value to
generator, transmit a signal
the corresponding input
equal to min. scale value to
of the instrument.
the corresponding input. For
input 4...20mA, give a
minimum value of 0mA.