Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 31
4. Operation

4.11. Access codes

The code 100 is for access to the calibration and programming of the instrument.
There are codes with other functions:
Code Action
401
Modification of the relay lag of the limits relays activation and deactivation.
Standard value: 2s. Programmable values: 0 to 99 s
413
Modification of the time for alarm activation when there is out of range reading.
Standard value: 60 s. Programmable values: 0 to 99 s/min/h
473
Modification of the delay time activation of the analogical output 4 to 20 mA after quitting Hold.
Standard value: 10 s. Programmable values: 0 to 99 s
497
Possibility to adjust the temperature for minimizing the effect of long cables.
22