Herunterladen Inhalt Inhalt Diese Seite drucken

Calibration Of Ph Electrode - Hach LANGE si628 P Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 31
4. Operation

4.7. Calibration of pH electrode

With the calibration the instrument adapts to the electrode and it's variations suffered over time.It is
recommended to calibrate periodically.
The process is carried out in three steps:
1. Selection of buffers temperature.
2. Calibration with buffer 7.
3. Calibration with a second buffer, pH 4 or 9.2.
If during calibration process an anomaly is detected
an error message will appear (see page 24).
Enter code 100
Enter buffers temperature with
The display shifts automatically when the reading is stable.
The second buffer used can be pH 4 or pH 9.2.
Theoretical value of the buffer at the calibration temperature.
Notes: During CALIBRATION instrument is in hold state:
• Contact relays are opened.
• Current output is frozen at 21 mA.
To quit the hold state it is necessary to return to MEASURE.
Press ESC to quit the calibration, from any screen, without
storing data.
For practical reasons, it is recommeded to select temperature
manually during CALIBRATION, even if you have an ATC
probe connected (Pt 1000).
si628 P_HACH_LANGE_Ed.0510.indd ANG:19
si628 P_HACH_LANGE_Ed.0510.indd ANG:19
keys
M E A S U R E
pH 7.34
25.2 °C
CODE 000
CALIBRATION
Put buffer
pH 7
25°C
CALIBRATING
pH 7.02
19°C
Put 2 nd buffer
CALIBRATING
pH 4.00
19°C
Electrode OK
MEASURE pH & °C
M E A S U R E
pH 7.37
25.2 °C
19
20/5/10 11:24:25
20/5/10 11:24:25

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis