4. Operation
4.5. Programming to measure ORP (redox). mV range
The instrument is supplied programmed to measure pH. To change the measuring unit (pH) to mV go to
Setting-up, after selecting the language, see page 14, or follow the diagram below until TYPE OF MEASURE.
MEASURE
375 mV
CODE 000
Enter code 100
Limit 1 y Limit 2.
Selection of the functioning val-
ues of the corresponding relays.
Input "ON" (closed contact).
Output "OFF" (open contact).
Selectable values between
± 1500 mV.
Output current.
Selection of mV values corre-
sponding to 4 mA and 20 mA.
Selectable values between
± 1500 mV.
Hold. Selection of the current
output when the instrument is in
Hold state: fixed at 21 mA or at
the mA corresponding to the last
measured value.
Language.
Selection of the working language:
English, Spanish, Italian or French.
Note: When programming, the
instrument is in hold state:
•
The contact relays are open.
•
The current output is frozen at
21 mA.
To quit the hold state it is neces-
sary to return to MEASURE.
The output current and relays will
activated after 10 s.
si628 P_HACH_LANGE_Ed.0510.indd ANG:17
si628 P_HACH_LANGE_Ed.0510.indd ANG:17
CALIBRATION
LIMIT 1 (L1)
LIMIT 2 (L2)
OUTPUT 4-20 mA
HOLD
LANGUAGE
TYPE OF MEASURE
MEASURE mV & °C
ON - - L1 - - OFF
-300 mV
-200 mV
Vary the underlined digit with
and accept with
ON - - L2 - - OFF
300 mV
200 mV
4 mA
20 mA
-500 mV +500 mV
> 21 mA
last
LANGUAGE > Español
Italiano
MEASURE OF: pH
> mV
Vary with
MEASURE
375 mV
25.2 °C
17
20/5/10 11:24:24
20/5/10 11:24:24