BLUETOOTH
CONFIGURATION

CONNECTION PROCEDURE

Symbol
off..............disconnected mode
Symbol
blinking.....advertising mode
Symbol
on...............connected mode
The following options can be selected to control the Bluetooth ® module.
Enable Bluetooth ® module (start advertising mode).
Disable Bluetooth ® module (terminate active connection).
Clear pairing information.
Display the MAC (Media Access Control) address.
Three Bluetooth ® profiles are available.
Profile without pairing (default).
Paired and secured profile.
Virtual keyboard mode (compatible with recent equipment without driver installation).
Note:
- Bluetooth ® pairing information is cleared when the profile is changed.
Connection:
1. Activate Bluetooth ® compatible software and hardware (Master : PC, Display Unit).
2. Start the instrument. By default the Bluetooth ® module is active and the instrument is available for
connection (advertising mode).
3 . If no connection is established during the advertisement period reactivate the Bluetooth ® module
using the
/
menu.
4. Instrument is ready to communicate (connected mode).
Only with paired profile:
Pairing with master is automatically done at first connection.
To connect the instrument to a new master (new pairing), pairing information on the instrument must be
cleared using the
/
®
BLUETOOTH
SPECIFICATIONS
Frequency Band
Modulation
Max. output power
Range
Battery life
Other specifications on the manufacturer's website.
menu.
Continuous : up to 2 months - Always connected with 4 values / sec.
Saver : up to 5 months - The instrument sends value only when the position has changed.
Blind/Push : up to 7 months - Value is sent from the instrument (button) or requested
from the computer.
2.4 GHz (2.402 - 2.480 GHz)
GFSK (Gaussian Frequency Shift Keying)
Class 3: 1mW (0dBm)
Open space: up to 15 m
Industrial environment: 1-5 m