Herunterladen Diese Seite drucken

Agilent Technologies TV 141 Navigator Vent Valve Bedienungshandbuch Seite 36

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4
Installation procedure
Installation
Once the kit is assembled, install it on the pump.
Remove the screw cap from the pump, then secure the valve to the
pump using a 16 mm wrench making sure to tighten the nut with a
torque equivalent to 2.5 Nm.
The following figure shows an example of installation.
Figure 3
To avoid damaging the pump, do not tighten the nut with a torque greater than
CAUTION!
2.5 Nm.
Upon completion of the mechanical installation, attach the
connection cable between the valve to the control unit, and the
connection cable between the electrovalve control unit to the pump's
controller. Use the velcrose provided to secure the controller box.
36/44
Valve
Control unit
Connection
cables
Installation
TV 141 Navigator Vent Valve User Manual / 87-900-911-01 (F)

Werbung

loading

Diese Anleitung auch für:

969-9834