Herunterladen Inhalt Inhalt Diese Seite drucken

Hach LANGE si629 P Bedienungsanleitung Seite 23

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 39
4. Operation
4.5. Programming (continuation)
Language. Select the
language among En-
glish, German,
Spanish, French,
Italian, Portuguese.
Relay programming with Proportional Control
LIMIT
Optimal value. pH or mV value at which the sample
must be mantained.
Tolerance. Zone around the optimal value, that is also
accepted as correct, and therefore does not activate
any control.
If only one limit relay is used, the tolerance can be
positive or negative, defining the limit as maximum or
minimum.
If 2 limit relays are configured then the tolerance is syme-
tric around the optimal value: if the tolerance is positive,
L1 is the maximum limit and L2 is the minimum limit. If the
tolerance is negative it is just the other way round.
si629 P_HACH_LANGE_Ed.0510.indd ANG:19
si629 P_HACH_LANGE_Ed.0510.indd ANG:19
LANGUAGE
Press
until MEASURE pH & ˚C is reached.
MEASURE pH & °C
Note: During Configuration the instrument is in Hold state:
MEASURE
• Limit and cleaning contact relays are opened. (Alarm OFF).
pH 7.34
25.2 °C
• Current output is frozen at the value programmed in Hold.
To quit the Hold state it is necessary to return to measure.
The Current output 4 to 20 and the Relays will activate
in 10 sec.
PROP. C. > Optimal v.
Tolerance
Proportional band
Cycle time
LANGUAGE >Spanish
Italian
Optimal value
pH
7.0
Tolerance
pH
+1.0
Proportional band
pH
01.0
Cycle time
2 s
Proportional band. Zone where the control is carried
out. In this area the control signal is active during a
time "proportional" to the distance between the reading
and the limits of the tolerance band.
Cycle time (2 to 99 sec.). It defines the relay activation
time in the proportional band.
Depending on the distance between the reading and
the limits of the tolerance band, the relay will be activa-
ted a percentage of the programmed cycle time.time
IMPORTANT: See Relays actuation in proportional
control, page 25.
19
20/5/10 13:00:52
20/5/10 13:00:52

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis