Herunterladen Diese Seite drucken

Start-Up - Agilent Technologies 4UHV Gebrauchsanleitung

Ion pump controller

Werbung

5
Technical Information

Start-Up

Start-Up

This chapter explains how to switch on and off the unit and the high
voltage channels. In the following chapters, all the functions of the
controller are explained in detail.
1
2
3
4
5
6
7
Figure 12
60/114
Put the main power switch situated on the front panel to OFF
Insert the power cable into the correct connector on the rear
panel
Connect the HV cables to the controller and to the ion pumps
Make sure that "Interlock" of HV cable is connected
To switch on the unit, put the main power switch to ON
Select Local for the Operating mode
Press the keys HV-ON/OFF keys and HV-x, where x is the number
of the channel you want to switch on.
Main power switch
Front Panel
Agilent 4UHV Ion Pump Controller User Manual
Key HV-ON/OFF
Keys HV-x
/
87-900-137-01 (D)

Werbung

loading