8.4.2 Object 2006h: External output function
About the External output a measuring system fault can be output. The following faults
can be chosen:
Index
Name
Data type
Access
Default
Value Description
0
1
2
3
4
5
6
Printed in the Federal Republic of Germany
04/17/2019
0x2006
Ext. Output Function
UNSIGNED8
rw
0
disabled:
Function switched off, following parameters without meaning. Depending
on "Object 2008h: " at the external output either "0"-level or "1"-level is
output.
Fail output temperature:
The switching output is set when the device temperature is above the limit.
A low range deviation has no influence on the measured value and should
therefore be regarded as a warning. The limit can be changed via
TRWinProg, default value 47 °C.
Fail output intensity:
The switching output is set, if an intensity value of smaller 8% is present, or
the laser beam is interrupted and leads to the failure value output.
Fail output hardware-failure:
The switching output is set, if an internal hardware error is detected and
leads to the failure value output.
Fail output every failure:
The switching output is set, if one of the errors, listed here, is active.
Speed-check:
The switching output is set, if the adjusted speed is exceeded. About the
default setting the speed-check is switched off. Configurability see chapter
8.6 on page 85.
Plausibility measurement-value:
The switching output is set, if the plausibility of the measured value cannot
be guaranteed. E.g. this is the case at a position jump if a second reflection
foil is held into the laser beam.
TR - ELE - BA - DGB - 0027 - 01
TR-Electronic GmbH 2018, All Rights Reserved
Page 83 of 99