Herunterladen Inhalt Inhalt Diese Seite drucken

General Information - Hameg HM8030-5 Handbuch

Vorschau ausblenden Andere Handbücher für HM8030-5:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

General information

The operator should not neglect to carefully
read the following instructions and those of the
mainframe HM8001, to avoid any operating errors
and to be fully acquainted with the module when
later in use.
After unpacking the module, check for any mecha-
nical damage or loose parts inside. Should there be
any transportation damage, inform the supplier
immediately and do not put the module into
operation.
This plug-in module is primarily intented for use in
conjunction with the Mainframe HM8001. When
incorporating it into other systems, the module
should only be operated with the specified supply
voltages.
Safety
This instrument has been designed and tested
in accordance with IEC Publication 1010-1, Safety
requirements for electrical equipment for
measurement, control, and laboratory use. It
corresponds as well to the the CENELEC
regulations EN 61010-1. All case and chassis parts
are connected to the safety earth conductor.
Corresponding to Safety Class 1 regulations (three-
conductor AC power cable). Without an isolating
transformer, the instruments power cable must be
plugged into an approved three-contact electrical
outlet, which meets International Electrotechnical
Commission (IEC) safety standards.
Warning!
Any interruption of the protective conductor
inside or outside the instrument or discon-
nection of the protective earth terminal is likely
to make the instrument dangerous. Intentional
interruption is prohibited.
The istrument must be disconnected and secured
against unintentional operation if there is any
suggestion that safe operation is not possible. This
may occur:
if the instrument has visible damage,
if the instrument has loose parts.
if the instrument does not function,
after long storage under unfavourable
circumstances (e.g. outdoors or in moist
environments),
after excessive transportation stress (e.g. in
poor packaging).
When removing or replacing the metal case, the
instrument must be completely disconnected from
the mains supply. If any measurement or calibration
procedures are unavoidable on the opened-up
instrument, these must only be carried out by qualified
personnel acquainted with the danger involved.
Symbols as Marked on Equipment
ATTENTION refer to manual.
DANGER High voltage.
14
Protective ground (earth) terminal.
Operating conditions
The ambient temperature range during operation
should be between +10°C and +40°C and should
not exceed −40°C or +70°C during transport or
storage. The operational position is optional,
however, the ventilation holes on the HM8001 and
on the plug-in modules must not be obstructed.
Warranty
Before being shipped, each plug-in module
must pass a 24 hour quality control test.
Provided the instrument has not undergone any
modifications Hameg warrants that all products of
its own manufacture conform to Hameg
specifications and are free from defects in material
and workmanship when used under normal
operating conditions and with the service conditions
for wich they were furnished.
The obligation of HAMEG hereunder shall expire
two (2) years after delivery and is limited to repairing,
or at its option, replacing without charge, any such
product which in Hamegs sole opinion proves to be
defective with the scope of this warranty.
This is Hamegs sole warranty with respect to the
products delivered hereunder. No statement,
representation, agreement or understanding, oral
or written, made by an agent, distributor,
representative or employee of, which is not
contained in thiss warranty will be binding upon
Hameg, unless made in writing and executed by an
authorized Hameg employee. Hameg makes no
other warranty of any kind whatsoever, expressed
or implied, and all implied warranties of
merchantibility and fitness for a particular use
which exceed the aforestated obligation are hereby
disclaimed by Hameg be liable to buyer, in contract
or in tort, for any special, indirect, incidental or
consequential damages, espresses, losses or
delays however caused.
In case of any complaint, attach a tag to the
instrument with a description of the fault observed.
Please supply name and department, address and
telephone number to ensure rapid service.
The instrument should be returned in its original
packaging for maximum protection. We regret
that transportation damage due to poor packaging
is not covered by this warranty.
Maintenance
The most important characteristics of the
instruments should be periodically checked
according th the instructions provided in the
sections "Operational check and "Alignment
procedcure. To obtain the normal operating
temperature, the mainframe with inserted module
should be turned on at least 60 minutes before
starting the test. The specified alignment procedure
should be strictly observed. When removing the
case detach mains/line cord and any other
connected cables from case of the mainframe
HM8001. Remove both screws on rear panel and,
Änderungen vorbehalten / Subject to change without notice

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Hm 8030-5Hm8001-2

Inhaltsverzeichnis