2. If the test does not start after applying the sample:
POSSIBLE CAUSE
Insufficient blood sample
Defective test strip
Sample applied after automatic switch-
off (3 minutes after last user action)
Defective meter
3. If the control solution testing result is out of range:
POSSIBLE CAUSE
Error in performing the test
Control solution vial was poorly shaken
Expired or contaminated control
solution
Control solution that is too warm or
too cold
Defective test strip
Meter malfunction
Improper working of meter and test
strip
58
WHAT TO DO
Repeat the test using a new test strip
with larger volume of blood sample
Repeat the test with a new test strip
Apply sample only when flashing " "
appears on the display
Please contact customer services
WHAT TO DO
Read instructions thoroughly and
repeat the test again
Shake the control solution vigorously
and repeat the test again
Check the expiry date of the control
solution
Control solution, meter, and test strips
should be at room temperature 20°C to
25°C before testing
Repeat the test with a new test strip
Please contact customer services
Please contact customer services

SYMBOL INFORMATION

SYMBOL
In vitro diagnostic medical device
Consult instructions for use
Temperature limit
Serial number
Authorized representative in the European
Humidity limitation
0123
Catalogue number
Do not discard in household waste
REFERENT
Use-by date
Batch code
Manufacturer
Community
Caution
CE mark
59