NOTE:
To avoid contaminating the control solution, do not directly apply control solu-
tion onto a strip.
Your device will tag this measurement as "QC" test automatically.
3. Read and compare the result
After counting down to 0, the control so-
lution test result will appear on the display.
Compare this result with the range prin-
ted on the test strip vial and it should fall
within this range. If not, please read the
instructions again and repeat the control solu-
tion test.
With "QC" displayed, the meter will store your
test result in memory under "QC".
NOTE:
The control solution range printed on the test strip vial is for control solution use
only. It is not a recommended range for your blood test level.
See the MAINTENANCE section for important information about your control
solutions.
46
f
GLU
GLU
Normal

TESTING WITH BLOOD SAMPLE

WARNING:
To reduce the chance of infection:
Never share a lancet or the lancing device.
Always use a new, sterile lancet. Lancets are for single use only.
Avoid getting hand lotion, oils, dirt, or debris in or on the lancets and the lancing
device.

Preparing the Lancing Device for Blood Testing

Please follow the instructions in the lancing device insert for collecting a blood sam-
ple.

Preparing the Puncture Site

Stimulating blood perfusion by rubbing the puncture site before blood extraction
has a significant influence on the accuracy of the glucose value obtained.
Blood from a site that has not been rubbed exhibits a measurably different glucose
concentration than blood from the finger. If the puncture site was rubbed prior to
blood extraction, the difference is significantly reduced.
Please follow the suggestions below before obtaining a drop of blood:
Wash and dry your hands before starting.
Select the puncture site at fingertips.
Rub the puncture site for about 20 seconds before punction.
Clean the puncture site using cotton moistened with 70% alcohol and let it air dry.
Fingertip testing
Press the lancing device's tip firmly against the lower
side of your fingertip. Press the release button to
prick your finger, then a click indicates that the punc-
ture is complete.