Herunterladen Diese Seite drucken

Mounting The Unit - Welch Allyn SureTemp Plus series Handbuch

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 119
2.
Connect the appropriate removable probe well to the thermometer. Note that
the cut-out in the probe well fits over the probe cord strain relief. Ensure that
the removable probe well "snaps" into position and is completely seated to
the thermometer. Place the probe into the well.
3.
Insert the box of Welch Allyn disposable probe covers into the probe cover
storage holder. To open the box of probe covers, remove and discard the tear-
away corner. You will see the ends of the probe covers in the box.

MOUNTING THE UNIT

Mount the wall-mount bracket in a location such as an interior wall or wall divider
so that the thermometer remains at ambient room temperature. Mounting the
unit on an uninsulated exterior wall, in an area directly hit by sunlight, above heat
radiators or powered equipment (including computers and lights), or below air
conditioning outlets is not recommended.
Proper mounting of the unit will prevent tangling of the probe cord with other
equipment at the patient's bedside.
1-8
1.
Mount the instrument so
that all other objects are
no closer than 4 inches
(10.25 cm) from the sides
and bottom of the unit.
2.
Mount the wall mount a
maximum of 48 inches
(122 cm) from the floor.
3.
Install the wall mount to
allow at least 12 inches
30.5 cm) clearance on top
of the instrument. This
provides ease of
removing and installing
the probe and probe well
or installing probe covers
onto the probe.

Werbung

loading

Diese Anleitung auch für:

Suretemp plus 690Suretemp plus 692