Herunterladen Inhalt Inhalt Diese Seite drucken

Riester RBP-100 Bedienungsanleitung Seite 59

Digitales blutdruckmessgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
 
 
 
Recommended  separation  distances  between    
portable  and  mobile  RF  communications  equipment  and  the  No  touch  Infrared  Body  Thermometer.  
The  RBP-­100  Blood  Pressure  Monitor  and  the  RBP-­100  USB  Blood  Pressure  Monitor  is  intended  for  use  in  
an  electromagnetic  environment  in  which  radiated  RF  disturbances  are  controlled.  The  customer  or  the  user  
of  the  RBP-­100  Blood  Pressure  Monitor  and  the  RBP-­100  USB  Blood  Pressure  Monitor  can  help  prevent  
electromagnetic  interference  by  maintaining  a  minimum  distance  between  portable  and  mobile  RF  
communications  equipment  (transmitters)  and  the  RBP-­100  Blood  Pressure  Monitor  and  the  RBP-­100  USB  
Blood  Pressure  Monitor  as  recommended  below,  according  to  the  maximum  output  power  of  the  
communications  equipment.  
Rated  maximum  
output  power  of  
150  KHz  to  80  MHz  
transmitter  
(W)  
0.01  
0.1  
1  
10  
100  
For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended  separation  distance  
d  inmetres  (m)  can  be  estimated  using  the  equation  applicable  to  the  frequency  of  the  transmitter,  where  P  
is  themaximum  output  power  rating  of  the  transmitter  in  watts  (W)  according  to  the  transmitter  manufacturer.  
NOTE  1  At  80  MHz  and  800  MHz,  the  separation  distance  for  the  higher  frequency  range  applies.  
NOTE  2  These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by  
absorption  andreflection  from  structures,  objects  and  people.  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Separation  distance  according  to  frequency  of  transmitter  
(m)  
80  MHz  to  800  MHz  
 
 
0.12  
0.12  
0.38  
0.38  
1.2  
1.2  
3.8  
3.8  
12  
12  
800  MHz  to  2.7  GHz  
 
 
0.23  
0.73  
2.3  
7.3  
23  
Seite 22 von 23